Skip to main content
Help

Revision to Parker Nitronox FM-1308

Kristen Gismondi

[flag|biomed]
-[flag|student_owned]
[title|Parker Nitronox FM-1308 Repair]
-[summary]Repair guide and service information for the Parker Nitronox FM-1308. This medical device provide a fast, safe, effective pain relief for patients in a variety of medical setting.[/summary]
+[summary]Repair guide and service information for the Parker Nitronox FM-1308. This medical device provides a fast, safe, and effective pain relief for patients in a variety of medical settings.[/summary]
[summary_image|2092412]
[document|14611|display=list]
-== ***Background and identification:*** ==
+== ***Background and Identification:*** ==
-The Parker Nitronox FM-1308 is one of the models of Nitronox product. It is an anesthesia system that provide a safe and fast pain relief for patients during the modern procedure. This device is designed for both children and adults. It is equipped with single patients breathing circuits and masks. It preset the N2O/O2 mixture to eliminating the risk of delivering hypoxic mixture.
+The Parker Nitronox FM-1308 is an anesthesia system that provides safe and fast pain relief for patients during medical procedures. This device is designed for both children and adults. It is equipped with single patient breathing circuits and masks. It has a preset with an N2O/O2 mixture to eliminate the risk of delivering hypoxic mixture.
-This company was founded in 1968, they has specialized in the development of medical device that measure and control the gases and liquid. The Nitronox series is one their product series. However, this Nitronox FM-1308 has been discontinued for a long time.
+This company was founded in 1968 and specialized in the development of medical devices that measure and control gases and liquid. The Nitronox FM-1308 has been discontinued for a long time.
-Nitronox has many model after FM-1308. It constantly updating until the Nitronox Plus comes out. The Nitronox Plus could predict the patients’ demand and provide three different models allow patients to choose for themselves.
+Nitronox has many models after the FM-1308 including the Nitronox Plus.
-== ***Technical specification:*** ==
+== ***Technical Specifications:*** ==
-***Nitronox Specifications***
-
-***Nitronox Housing Dimensions (H x W x D):***   9” x 5 ½” x 5”
-
-***Portable Stand Dimensions***
-
-***Overall Height Stand:***                   49” max – 41” min
-
-***Stand Width at Base:***                   22”
-
-***Weight:***                              8.2 pounds
-
-***Mixed Gas Outlet Connection:***        Black hose to quick disconnect to Demand Valve
-
-***Oxygen Inlet***:                   DISS ½” OD (Diameter Indexed Safety System) Green Hose
-
-***Nitrous Oxide Inlet:***              DISS 3/8” OD (Diameter Indexed Safety System) Blue Hose
-
-***Inlet Pressure (Each gas):***           50 – 55 psi (adjusted at E-Stand); 40 – 65 psi (central gas supply; O2 input pressure cannot exceed N2O input pressure by 15 psi or higher
-
-***Instantaneous Flow Capability:***       114 LPM Maximum
-
-***Delivered Mixture Concentration:***    50% N2O and 50% O2 (+/- 5 percentage points O2; at typical deep breathing)
-
-***Nitronox Scavenger Vacuum Connection:***     3/8” ID vacuum hose on 3/8” hose barb
-
-***Scavenger Hose***:                            19 mm ID
-
-***Vacuum Flow (dependent on vacuum source):***       30 – 35 LPM Maximum (ball valve lever in full on position)
-
-***Operating Temperature:***                   40ºF (5ºC) minimum; 140ºF (60ºC) maximum
+* ***Nitronox Housing Dimensions (H x W x D):***   9” x 5 ½” x 5”
+* ***Portable Stand Dimensions***
+** ***Overall Height Stand:*** 49” max – 41” min
+** ***Stand Width at Base:***  22”
+* ***Weight:*** 8.2 pounds
+* ***Mixed Gas Outlet Connection:*** Black hose to quick disconnect to Demand Valve
+* ***Oxygen Inlet***: DISS ½” OD (Diameter Indexed Safety System) Green Hose
+* ***Nitrous Oxide Inlet:*** DISS 3/8” OD (Diameter Indexed Safety System) Blue Hose
+* ***Inlet Pressure (Each gas):*** 50 – 55 psi (adjusted at E-Stand); 40 – 65 psi (central gas supply; O2 input pressure cannot exceed N2O input pressure by 15 psi or higher
+* ***Instantaneous Flow Capability:*** 114 LPM Maximum
+* ***Delivered Mixture Concentration:*** 50% N2O and 50% O2 (+/- 5 percentage points O2; at typical deep breathing)
+* ***Nitronox Scavenger Vacuum Connection:*** 3/8” ID vacuum hose on 3/8” hose barb
+* ***Scavenger Hose***: 19 mm ID
+* ***Vacuum Flow (dependent on vacuum source):*** 30 – 35 LPM Maximum (ball valve lever in full on position)
+* ***Operating Temperature:*** 40ºF (5ºC) minimum; 140ºF (60ºC) maximum
***Source:***