Puritan Bennett 560 Exhalation Block Replacement

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.
  1. Puritan Bennett 560 Exhalation Block Replacement, Exhalation Block: step 1, image 1 of 1
    • Locate the exhalation block on the front of the ventilator.

  2. Puritan Bennett 560 Exhalation Block Replacement: step 2, image 1 of 1
    • Tilt the machine up and unscrew the knob that holds the exhalation block in place.

    • This knob can be untwisted by hand, no tools required.

  3. Puritan Bennett 560 Exhalation Block Replacement: step 3, image 1 of 1
    • Pinch the "From Patient" fitting and slide the exhalation block away from the ventilator.

  4. Puritan Bennett 560 Exhalation Block Replacement: step 4, image 1 of 1
    • After replacing or cleaning the block, perform the exhalation flow sensor calibration.

    • See 6.7.3.3 in the Service Manual for instructions on calibration.

Conclusion

To reassemble your device, follow these instructions in reverse order.