Low oxygen flow drops when plug in tubing
Indicator ball drops when plug in tubing on M10 concencentrator
Is this a good question?
Indicator ball drops when plug in tubing on M10 concencentrator
Is this a good question?
I'm my case, it was because the tubing had become clogged with water from the "humidifier". Also, we use a very long line and that reduces oxygen flow a lot. Watch that tubing length, it matters.
Was this answer helpful?
you have a blockage on the tubing or the bubble humidifier is damaged.
Was this answer helpful?